Introduction
Fiber-optic evanescent wave biosensors FOB Series are constructed
based on the principle of evanescent wave while light propagates in an
optical fiber. When the target biomolecules in question are
specifically bound to the capture component immobilized on the fiber
core surface, the fluorophores labeled to the target biomolecules
become attached within the evanescent zone of the core surface at the
same time. Excitation of fluorophore in the evanescent zone occurs
when light is launched into the fiber. Fluorescent emission at a
longer wavelength is trapped in the fiber and is then coupled back
into the fiber. The fluorescent flux relates to the concentration of
target biomolecules.
This type of biosensors has several
advantages over other fluorescence-based sensors. The evanescent wave
penetrates essentially only a fraction of a wavelength into the sample
medium and thus excites fluorescence primarily from the fluorescently
labeled analyte complexes that have been bound to the surface through
specific antibody-antigen recognition reaction (i.e., immunoassay) and
hybridization of nucleic acid. Not only does this reduce the
fluorescence from the sample or unbound fluorescent reagents, but the
signal obtained is directly related to the binding kinetics of the
detection interaction, i.e., a rapid real-time response can be
observed without complicated washing or separation steps, and field
detection of a variety of analytes can be performed.
Features
|
Model |
Performances |
|
FOB-2 |
l
5 fiber probes allow for simultaneous detection of 5 agents within one sample
l Built-in fluidic system including
computer-controlled peristaltic pump and valves, allows
automated sucking and exhausting of sample, buffer and
fluorophore-labeled secondary antibody
l Built-in multi-function DAS card controls the test
procedure, and processes and display and stores result data
l A warming system is used to keep the sample at a
constant temperature of 37°Cor45°C
l Low sample volume, less than 2 ml |
|
FOB-3 |
l
Single fiber probe
l Built-in multi-function DAS card connects to
either notebook or desktop computer
l A warming system is used to keep the sample at a
constant temperature of 37°Cor45°C
l Rechargeable battery powered for 8 hrs continuous
operation normally
l Dimension (mm): 235(W)´300(D)´120(H)
l Low sample volume, less than 250
ml |
Specifications
1) Light source: 635-nm laser diode to excite Cy5 fluorophore.
Other light sources are available for excitation of other fluorophores
2) Sensing element: fiber of fused silica, or injection-molded polystyrene fiber
3) High specificity and sensitivity lower than 0.01nmol/l for pure Cy5 solution
4) User-friendly and ease of use software involved
5) Real-time binding measurement can be performed
Fiber-optic Evanescent Wave Biosensor FOB-2

Fiber-optic Evanescent Wave Biosensor FOB-3
 |